Daily maintenance and maintenance of pressure steam sterilizer


1. Wipe the surface of the sterilizer door and instrument daily, and clean the floor between the sterilization chamber at least once.
2. Clean the impurities in the strainer at the drain port of the sterilizer daily to prevent impurities from entering the exhaust pipe during sterilizer operation.
3. Check whether the door seal of the sterilizer is smooth, intact, and not removed or damaged before daily operation.
4. Check the accuracy of the meter pointer every day, observe whether the temperature meter and pressure instrument pointer is in “0” position after the sterilizer operation is stopped; observe whether the steam, water and other medium pipelines and valve parts have leaked; Bacteria operation indicator is in good condition; once it is found that there is a problem with the above components, the sterilizer should not be used. It should be used after maintenance.
5. Clean the wipe inside the sterilizer weekly and clean thoroughly.
6. Clean the exterior of the sterilization equipment quarterly to avoid dust accumulation and shorten the life of the air filter. Components should be kept out of contact with water. Once the wet water has dried, turn on the power supply.
7. Every quarter, check whether the outlets and connectors of each connection are loose and loose.
8. Clean the safety valve surface every 6 months.
9. Check the steam line filter once a week and record the results.

10. Each year, the pressure steam sterilizer is tested once a year, and safety valves, pressure gauges, and thermometers are validated at least once a year. The inspection results are recorded and retained. The air filter should be replaced regularly.